AVALIAÇÃO DA LIMPEZA EM INSTRUMENTAIS DE VIDEOLAPAROSCOPIA COM TESTE DE ADENOSINA TRIFOSFATO (ATP)
dc.contributor.advisor | Klafke, André | |
dc.creator | Oliveira, Luciana Lima de | |
dc.date.accessioned | 2023-09-21T12:49:03Z | |
dc.date.available | 2023-09-21T12:49:03Z | |
dc.date.created | 2023-09-21 | |
dc.date.issued | 2020-10-02 | |
dc.description.abstract | O objetivo deste trabalho é monitorar a limpeza de instrumentais de conformação complexa da bandeja de videolaparoscopia utilizando o teste de adenosina trifosfato (ATP) no Centro de Material e Esterilização do Hospital Nossa Senhora da Conceição (HNSC). Os objetivos específicos são: avaliar a redução percentual de unidades relativas de luz (RLU) no teste de ATP com a limpeza manual seguida de automatizada; mensurar o incremento do teste de ATP sobre a inspeção visual na avaliação de limpeza do material; propor um ponto de corte em RLU para definir “limpeza” dos instrumentos estudados; e avaliar a qualidade da inspeção visual como monitor de limpeza, em relação ao valor de RLU determinado após o processo de limpeza. Trata-se de um estudo do tipo transversal, analisando instrumentais antes e depois da limpeza e teste de concordância na inspeção visual. Foram analisadas 183 peças, e se observou uma redução de mais de 99% no valor de RLU. Quando utilizado o valor de corte de 100RLU, o teste de ATP se mostrou apenas 0,55% mais sensível que a inspeção visual. Ao reduzir o valor de corte conforme a amplitude interquartil da amostra, o teste de ATP foi 15,30% mais sensível. Os resultados encontrados demonstraram que o processo de limpeza está sendo efetivo na instituição estudada. Considerando o valor de RLU encontrado na pesquisa, sugere-se que se busque alcançar esse patamar de qualidade na prática de limpeza, e que o valor de corte de 100RLU não represente um valor compatível com o máximo de segurança que se pode oferecer ao paciente. Foi demonstrada fragilidade da inspeção visual como monitor de limpeza, uma vez que a subjetividade do avaliador pode interferir no resultado da avaliação – e a concordância entre os avaliadores foi fraca no estudo. Um monitor de objetividade numérica, como o teste de ATP, pode conferir mais segurança na prática avaliativa. | |
dc.identifier.uri | https://axis.ghc.com.br/handle/123456789/82 | |
dc.language.iso | Português | |
dc.local | Porto Alegre | |
dc.notes | The objective of this work is to monitor the cleaning of complex forming instruments on the videolaparoscopy tray using the triphosphate adenosine (ATP) test at the Material and Sterilization Center of Hospital Nossa Senhora da Conceição (HNSC). The defined objectives are: to evaluate the percentage reduction of units applied to light (RLU) in the ATP test with manual cleaning followed by automated cleaning; measure or increment the ATP test on a visual inspection in the assessment of cleanliness of the material; propose a cut-off point in RLU to define “cleanliness” of the studied instruments; and evaluate the quality of the visual inspection as a cleaning monitor, in relation to the RLU value determined after the cleaning process. This is a cross-sectional study, analyzing instruments before and after cleaning and test of concordance in visual inspection. 183 parts were analyzed, and a reduction of more than 99% in the RLU value was observed. When used or with a cut-off value of 100RLU, the ATP test shows only 0.55% more sensitive than visual inspection. When reducing the cutoff value according to the sample's interquartile range, the ATP test was 15.30% more sensitive. The results found demonstrated that the cleaning process is being effective in the studied institution. It is suggested that one seeks to reach the RLU value found in the research, and that the cut-off value of 100RLU does not represent a value compatible with the maximum safety that can be offered to the patient. Weakness of visual inspection as a cleaning monitor was demonstrated, since the subjectivity of the evaluator can interfere with the evaluation result - and an agreement between the evaluators was reduced in the study. A numerical objectivity monitor, such as the ATP test, can provide more security in evaluative practice. | |
dc.publisher | Grupo Hospitalar Conceição | |
dc.publisher.program | PROGRAMA DE PÓS-GRADUAÇÃO EM AVALIAÇÃO DE TECNOLOGIAS PARA O SUS - MESTRADO PROFISSIONAL EM AVALIAÇÃO E PRODUÇÃO DE TECNOLOGIAS PARA O SUS | |
dc.rights | Acesso Aberto | |
dc.rights | Attribution-NonCommercial-NoDerivs 3.0 Brazil | en |
dc.rights.uri | http://creativecommons.org/licenses/by-nc-nd/3.0/br/ | |
dc.subject | Saúde Pública | |
dc.subject | Brasil | |
dc.subject | Sistema Único de Saúde | |
dc.subject | Videolaparoscopia | |
dc.subject | Teste de Adenosina Trifosfato | |
dc.title | AVALIAÇÃO DA LIMPEZA EM INSTRUMENTAIS DE VIDEOLAPAROSCOPIA COM TESTE DE ADENOSINA TRIFOSFATO (ATP) | |
dc.title.alternative | EVALUATION OF CLEANING IN VIDEOLAPAROSCOPY INSTRUMENTS WITH ADENOSINE TRIPHOSPHATE (ATP) TEST | |
dc.type | Dissertação |